Test Modes & Ratings

Voltage Source

Output voltage:

± 100 mVpeak to ± 200 Vpeak

Max. output current

50 mApeak

 

Frequency Domain Spectroscopy (FDS)

Measurement current:

max. ± 50 mApeak

Frequency range:

5 μHz to 5 kHz

 

Time Domain Current Measurement (PDC)

Measurement current:

max. 10 mA

Frequency range:

20 µHz to 100 mHz

Input resistance:

2 kΩ

 

Combined Mode (FDS and PDC)

Frequency range:

20 μHz to 5 kHz

Time reduction:

Up to 75 % in comparison to exclusive FDS measurements

 

Capacitance, Dissipation Factor and Impedance

Tanδ range:

0 to 10

Capacitance range:

10 pF to 100 μF

Impedance range:

100 Ω to 20 TΩ

 

Accuracy

Time Domain Current Measurement (PDC)

Current accuracy:

0.5 % ± 1 pA


Phase Angle

Accuracy:

< 20 m°

Resolution:

< 0.6 m°


Dissipation Factor

Accuracy 1 @ 20 °C

for 1 mHz < f < 100 Hz

1 % + 3x10-4

for f < 1 mHz

2 % + 5x10-4

for f > 100 Hz

2 % + 5x10-4


Capacitance

Accuracy 2 @ 20 °C

0.5 % + 1 pF

 

1 Dissipation factor accuracy for capacitances > 100 pF
2 Capacitance accuracy for frequencies 100 mHz - 5 kHz

General

System Requirements

Operating system:

Windows 7 SP1, 8, 10 or higher;
32-bit and 64-bit

CPU:

Current Intel or AMD CPU

RAM

min. 2 GB

Interface

USB 2.0 or higher


AC Power Supply

For safety reasons only use grounded DRA power supply which is delivered with the SPECTANO 100.
DO NOT use any other power supply.


Environmental Conditions of Device (not accessories)

Operation temperature:

-10 °C to +55 °C / -31 °F to +131 °F

Storage temperature:

-10 °C to +65 °C / -31 °F to +149 °F

Relative humidity:

10 % to 95 % non-condensing

Air pressure:

70 kPa to 106 kPa

 

Mechanical Data

Dimensions:

260 x 50 x 256 mm / 10.25 x 2 x 10.5 inch (w x h x d)

Weight:

2.3 kG / 5 Ib

For more information on accessory technical data, check the accessories section.