SPECTANO 100 - Technical Data

SPECTANO 100 Technical Data

(208 KB)

Download

Voltage Source

Output voltage:± 100 mVpeak to ± 200 Vpeak
Max. output current:50 mApeak

 

Frequency Domain Spectroscopy (FDS)

Measurement current:max. ± 50 mApeak
Frequency range:5 μHz to 5 kHz

 

Time Domain Current Measurement (PDC)

Measurement current:max. 10 mA
Frequency range:20 µHz to 100 mHz
Input resistance:2 kΩ

 

Combined Mode (FDS and PDC)

Frequency range:20 μHz to 5 kHz
Time reduction:Up to 75 % in comparison to exclusive FDS measurements

 

Capacitance, Dissipation Factor and Impedance

Tanδ range:0 to 10
Capacitance range:10 pF to 100 μF
Impedance range:100 Ω to 20 TΩ

Time Domain Current Measurement (PDC)

Current accuracy:0.5 % ± 1 pA


Phase Angle

Accuracy:< 20 m°
Resolution:< 0.6 m°


Dissipation Factor

Accuracy 1 @ 20 °Cfor 1 mHz < f < 100 Hz1 % + 3x10-4
 for f < 1 mHz2 % + 5x10-4
 for f > 100 Hz2 % + 5x10-4


Capacitance

Accuracy 2 @ 20 °C0.5 % + 1 pF

 

1 Dissipation factor accuracy for capacitances > 100 pF
2 Capacitance accuracy for frequencies 100 mHz - 5 kHz

General

System Requirements

Operating system:Windows 7 SP1, 8, 10 or higher;
32-bit and 64-bit
CPU:

Current Intel or AMD CPU

RAMmin. 2 GB
InterfaceUSB 2.0 or higher


AC Power Supply
For safety reasons only use grounded DRA power supply which is delivered with the SPECTANO 100.
DO NOT use any other power supply.
 

Environmental Conditions of Device (not accessories)

Operation temperature:-10 °C to +55 °C / -31 °F to +131 °F
Storage temperature:

-10 °C to +65 °C / -31 °F to +149 °F

Relative humidity:10 % to 95 % non-condensing
Air pressure:70 kPa to 106 kPa

 

Mechanical Data

Dimensions:260 x 50 x 256 mm / 10.25 x 2 x 10.5 inch (w x h x d)
Weight:

2.3 kG / 5 Ib

For more information on accessory technical data, check the accessories section.

For more information on the SPECTANO 100 series, check the product description and knowledge & application section.